Title | In vitro assessment of long-term reliability of low-cost μECoG arrays. |
Publication Type | Conference Paper |
Year of Publication | 2016 |
Authors | K Palopoli-Trojani, V Woods,, M Trumpis, and J Viventi |
Conference Name | Annual International Conference of the Ieee Engineering in Medicine and Biology Society. Ieee Engineering in Medicine and Biology Society. Annual International Conference |
Date Published | 08/2016 |
Abstract | Micro-electrocorticographic (μECοG) electrode arrays provide a minimally invasive, high-resolution neural interface with broad cortical coverage. Previously, we fabricated μECoG arrays at a lower cost than commercially available devices using low-cost industrial processes [1], [2]. Here, we report the in vitro electrical performance of five μECoG designs undergoing an accelerated aging protocol. The impedance and yield of the μECoG arrays were tracked over time. The equivalent lifetime at 37°C depended on the manufacturer and material stack-up, and ranged between 30 and greater than 760 days (ongoing). The main failure modes of these devices were delamination at the site of the electrode contact and broken traces due to metal dissolution. Based on these in vitro results, we offer several recommendations for μECoG designs suitable for chronic implantation. |
DOI | 10.1109/embc.2016.7591728 |